SYRINGE 0.3ML 31GA 6MM 10BAG 500 AP
Report
- Report Number
- 1920898-2021-00468
- Event Type
- Malfunction
- Date Received
- April 27, 2021
- Date of Event
- March 30, 2021
- Report Date
- April 13, 2021
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT FOR SCALE MARKING DEFECTIVE ON LOT # 0132186. INVESTIGATION SUMMARY: CUSTOMER RETURNED (5) 3/10CC, 6MM, 31G SYRINGES IN AN OPEN POLY BAG FROM LOT # 0132186. CUSTOMER STATES THAT THE PRINTING OF SYRINGE SCALE MARK WAS SOMETHING WRONG. ALL RETURNED SYRINGES WERE TESTING USING THE PLUG GAUGE AND ALL SCALE MARKING PLACEMENTS FELL WITHIN SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 0132186 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. BASED ON THE SAMPLES AND/OR PHOTO(S) RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CAPA/SA IS REQUIRED AT THIS TIME.
IT WAS REPORTED THAT 5 SYRINGE 0.3ML 31GA 6MM 10BAG 500 AP HAD SCALE MARKING ISSUES BEFORE USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE PATIENT STATED THAT THERE WAS A LIQUID CONDENSATION AT THE TIP OF THE NEEDLE AND THE PRINTING OF SYRINGE SCALE MARK WAS SOMETHING WRONG."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631532 | SYRINGE 0.3ML 31GA 6MM 10BAG 500 AP | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 0132186 |