Narratives
Additional Manufacturer Narrative · 1
A GE REPRESENTATIVE EVALUATED THE SYSTEM, AND REPLACED THE S-RAM CARD. SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
CUSTOMER REPORTED THE SYSTEM IS DISPLAYING A CHECKSUM ERROR. NO PATIENT INJURY WAS REPORTED.