60 IN NON-DEHP MICROBORE EXTENSION SET
Report
- Report Number
- 9616066-2021-50819
- Event Type
- Malfunction
- Date Received
- April 26, 2021
- Date of Event
- February 8, 2021
- Report Date
- April 16, 2021
- Manufacturer
- SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 50885403236257
- PMA / PMN Number
- K051499
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE INITIAL REPORTER ALSO NOTIFIED THE FDA VIA MEDWATCH # (B)(4). INVESTIGATION SUMMARY: THE CUSTOMER REPORTED THAT THE END CONNECTOR BROKE OFF COMPLETELY, AND RETURNED THE USED SAMPLE, AND AN UNOPENED SAMPLE OF THE SAME LOT. THE UNUSED SAMPLE WAS OPENED AND TESTED FOR ANY DEFECTS, AND NONE WERE FOUND. THE SAMPLE WAS CLEARLY SEPARATED AT THE MALE LUER, AND THE COMPLAINT IS VERIFIED. THE USED SAMPLE WAS OBSERVED UNDER THE MICROSCOPE TO CHECK FOR DEFECTS. A DEVICE HISTORY RECORD REVIEW FOR MODEL ME2020 LOT NUMBER 19125802 WAS PERFORMED. THE SEARCH SHOWED THAT A TOTAL OF 14,403 UNITS IN 1 LOT NUMBER WAS BUILT ON 09DEC2019. THERE WERE NO QUALITY NOTIFICATIONS ISSUED FOR THE FAILURE MODE REPORTED BY THE CUSTOMER DURING THE PRODUCTION BUILD OF THIS SET. THE ROOT CAUSE IS INSUFFICIENT SOLVENT APPLIED DURING ASSEMBLY. THIS INCIDENT HAS BEEN ADDED TO OUR DATABASE OF REPORTED INCIDENTS. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT THE 60 IN NON-DEHP MICROBORE EXTENSION SET BROKE OFF FROM THE TUBING AFTER THE INFUSION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE END CONNECTOR OF THE CAREFUSION 303, INC. MAXGUARD EXTENSION SET BROKE OFF COMPLETELY FROM THE TUBING AFTER THE INFUSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 618755 | 60 IN NON-DEHP MICROBORE EXTENSION SET | INTRAVASCULAR ADMINISTRATION SET | FPA | SISTEMAS MEDICOS ALARIS, S.A. DE C.V. | ME2020 | 19125802 | 50885403236257 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |