FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE VOLUMETRIC INFUSION PUMP
MDR report key: 1172935
·
Received September 25, 2008
Report
- Report Number
- 6000001-2007-02716
- Event Type
- Malfunction
- Date Received
- September 25, 2008
- Date of Event
- January 1, 2007
- Report Date
- January 3, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 24 2007. EVALUATION SUMMARY:THE CONDITION OF FAIL CODE 812:02 WAS CONFIRMED DURING PRODUCT EVALUATION. FAIL CODE 812:02 WAS OBSERVED IN THE EVENT HISTORY AND WAS CAUSED BY A NON OPERATIONAL PUMP HEAD MODULE. THE PUMP HEAD MODULE WAS REPLACED.
Description of Event or Problem · 1
THE FACILITY REPRESENTATIVE REPORTED AN INFUSION PUMP WITH FAILURE CODE 812:02 DURING BIOMED TESTING. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |