FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 29GA 1/2IN

MDR report key: 11728789 · Received April 26, 2021

Report

Report Number
1920898-2021-00462
Event Type
Malfunction
Date Received
April 26, 2021
Date of Event
March 30, 2021
Report Date
April 2, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 5TH COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE HENCE THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED AT THIS TIME. INVESTIGATION CONCLUSION: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 5TH RELATED COMPLAINT FOR NEEDLE HUB SEPARATES ON LOT # 0160988. A REVIEW OF RISK MANAGEMENT (B)(4) REVISION 14 INDICATES THAT THE POTENTIAL RISK OF THIS SPECIFIC REPORTED INCIDENT (SYRINGE, NEEDLE HUB SEPARATES) WAS CAPTURED AND ADDRESSED. INVESTIGATION SUMMARY: CUSTOMER RETURNED (1) 3/10CC, 12.7MM, 29G SYRINGE IN AN OPEN POLY BAG FROM LOT # 0160988. CUSTOMER STATES THAT THERE IS NEEDLE/SHIED SEPARATION FROM THE BARREL. THE RETURNED SYRINGE WAS TESTED AND THE HUB ASSEMBLY DID NOT SEPARATE FROM THE BARREL WHEN THE SHIELD WAS REMOVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 0160988. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BASED ON THE SAMPLES AND/OR PHOTO(S) RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYRINGE 0.3ML 29GA 1/2IN EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER REPORTED ABOUT NEEDLE/SHIED SEPARATION FROM THE BARREL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620151 SYRINGE 0.3ML 29GA 1/2IN SYRINGE FMF BD MEDICAL - DIABETES CARE 0160988

Patients

Seq Age Sex Outcome Treatment
1