FDA Adverse Event Injury Summary report: N

INVACARE RELIANT 450

MDR report key: 1172825 · Received July 25, 2008

Report

Report Number
1172825
Event Type
Injury
Date Received
July 25, 2008
Date of Event
July 11, 2008
Report Date
July 24, 2008
Manufacturer
INVACARE LINAK U.S. INC.
Product Code
FSA
Product Problem
Yes
Report Source
User Facility report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

AN INVACARE RELIANT 450 FULL BODY SLING LIFT WAS BEING USED TO TRANSFER A NON-WEIGHT BEARING PT FROM THE BED TO CHAIR. THE BOLT THAT ATTACHES THE SLING HARDWARE TO THE MECHANICAL LIFT ARM HAD WORKED ITS WAY OUT, AND THE PT AND SLING ASSEMBLY DETACHED, AND BOTH FELL TO THE FLOOR. THE NUT THAT ATTACHED THE SLING ASSEMBLY HAD FALLEN OFF. NO COTTER KEY ON SAFETY WIRE TO KEEP IT SECURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INVACARE RELIANT 450 FULL BODY PATIENT LIFT FSA INVACARE LINAK U.S. INC. RPL-450-1

Patients

Seq Age Sex Outcome Treatment
1 78 YR Hospitalization| L| O