BD FACS SAMPLE PREP ASSISTANT I
Report
- Report Number
- 2916837-2008-00007
- Event Type
- Other
- Date Received
- September 12, 2008
- Date of Event
- August 15, 2008
- Report Date
- August 15, 2008
- Manufacturer
- TECAN US INC.
- Product Code
- GKZ
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
BD BIOSCIENCES WILL SEND A CUSTOMER LETTER AND USER BULLETIN IN THE FIELD CORRECTIVE ACTION KIT TO ALL BD FACS SAMPLE PREP I CUSTOMERS ADVISING THEM TO FOLLOW THE NEW PROBE REPLACEMENT PROCESS WHEN CHANGING THE PROBE AS A PART OF THEIR PREVENTATIVE MAINTENANCE PROGRAM. AN INTERNAL SERVICE BULLETIN WILL ALSO BE SENT TO ALL BD BIOSCIENCES FIELD SERVICE ENGINEERS ADVISING THEM TO FOLLOW THE NEW PROCESS WHEN CHANGING THE SPA PROBE TO REDUCE THE RISK.
A BD FIELD SERVICE ENGINEER WAS CONDUCTING A PREVENTATIVE MAINTENANCE PROCEDURE ON THE BD SAMPLE PREP ASSISTANT I. HE WAS FOLLOWING RECOMMENDED SAFETY PRECAUTIONS WEARING GLOVES AND A SMOCK. THE INSTRUMENT HAD GONE THROUGH A CLEANING PROCEDURE USING BD FACSCLEAN WHICH CONTAINS 10% BLEACH 2 HOURS PRIOR. THE CUSTOMER ASKED HIM TO REPLACE THE PROBE. HE FOLLOWED THE PROBE REPLACEMENT PROCEDURE BY LOOSENING THE SET SCREW. HE ENCOUNTERED SOME DIFFICULTY REMOVING THE PROBE WITH THE PLASTIC CASE ENCAPSULATING THE PROBE NEEDLE WHICH WAS SLIPPING OFF. HE THEN REMOVED THE PLASTIC CASING AND CONTINUED USING PLIERS TO PULL THE PROBE. WITH THE PLIERS HE PULLED THE PROBE STRAIGHT DOWN WITH HIS RIGHT HAND WHILE HOLDING THE SHAFT STEADY WITH HIS LEFT. DUE TO THE REQUIRED FORCE TO BREAK THE PROBE LOOSE, THE TIP OF THE PROBE HIT THE BOTTOM OF THE UNIT CAUSING THE BACK END OF THE BLUNT PROBE TO BOUNCE UP AND STRIKE HIS LEFT THUMB. A NEEDLE STICK OCCURRED WHICH RESULTED IN A SMALL PUNCTURE AND THE ENGINEER NOTICED A SMALL DROP OF HIS BLOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD FACS SAMPLE PREP ASSISTANT I | SPA I | GKZ | TECAN US INC. | 334080 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |