FDA Adverse Event Malfunction Summary report: N

PULSE GEN MODEL 101

MDR report key: 1172607 · Received September 19, 2008

Report

Report Number
1644487-2008-02254
Event Type
Malfunction
Date Received
September 19, 2008
Date of Event
January 15, 2002
Report Date
August 22, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE REPORTER INDICATED THAT AN EXPLANTED VNS GENERATOR, WHICH WAS CONFIRMED TO BE AT END OF SERVICE (EOS), EXHIBITED AN "OUT-OF-LIMIT (HIGH) SUPPLY CURRENT 1MA" DURING PRODUCT ANALYSIS. LATER, THIS ANALYSIS REVEALED THAT THE EVENT WAS DUE TO A SELECTABLE VALUE R35 RESISTOR. THOUGH THE OUT-OF-LIMIT SUPPLY CURRENT COULD HAVE POTENTIALLY BEEN A CONTRIBUTING FACTOR TO THE EOS OF THE DEVICE, THE REMAINING BATTERY LONGEVITY WAS CALCULATED TO BE -2.38 CONFIRMING THAT EOS WAS AN EXPECTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PULSE GEN MODEL 101 LYJ CYBERONICS, INC. 101

Patients

Seq Age Sex Outcome Treatment
1 24 YR