FDA Adverse Event Malfunction Summary report: N

PROGRAMMING WAND

MDR report key: 1172602 · Received September 19, 2008

Report

Report Number
1644487-2008-02262
Event Type
Malfunction
Date Received
September 19, 2008
Date of Event
August 22, 2008
Report Date
August 22, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

OTHER - DEVICE FAILURE WAS CONFIRMED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

THE REPORTER INDICATED THAT SHE WAS UNABLE TO USE HER VNS PROGRAMMING WAND TO INTERROGATE A PATIENT'S VNS DEVICE AND THAT THERE WAS A "BIG CUT" IN THE SERIAL ADAPTER CORD. THE PROGRAMMING WAND WAS RETURNED TO THE MANUFACTURER AND THE REPORTED EVENTS WERE CONFIRMED. PRODUCT ANALYSIS REVEALED THAT THE EVENT WAS RELATED TO AN OPEN CONDUCTOR AND PHYSICAL DAMAGE TO THE SERIAL DATE CABLE. A KNOWN GOOD BENCH SERIAL DATA CABLE AND BATTERY WAS INSTALLED, AND ALL COMMUNICATION ERRORS WERE RESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING WAND UNKNOWN LYJ CYBERONICS, INC. 201

Patients

Seq Age Sex Outcome Treatment
1