DIMENSION CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2008-00105
- Event Type
- Malfunction
- Date Received
- September 19, 2008
- Date of Event
- August 30, 2008
- Report Date
- August 31, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- JJE
- PMA / PMN Number
- K944093
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A SIEMENS HEALTHCARE DIAGNOSTICS INC. FIELD SERVICE REPRESENTATIVE (FSE) WAS DISPATCHED TO THE ACCOUNT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED TROPONIN I RESULT LACK OF REAGENT DELIVERY DUE TO SYRINGE AND TUBING NOT CONNECTED PROPERLY. THE FSE CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY DEPRESSED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A POSITIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | JJE | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | RXLMAXRH | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |