FDA Adverse Event Malfunction Summary report: N

DIMENSION CLINICAL CHEMISTRY SYSTEM

MDR report key: 1172386 · Received September 19, 2008

Report

Report Number
1226181-2008-00105
Event Type
Malfunction
Date Received
September 19, 2008
Date of Event
August 30, 2008
Report Date
August 31, 2008
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Product Code
JJE
PMA / PMN Number
K944093
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS HEALTHCARE DIAGNOSTICS INC. FIELD SERVICE REPRESENTATIVE (FSE) WAS DISPATCHED TO THE ACCOUNT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED TROPONIN I RESULT LACK OF REAGENT DELIVERY DUE TO SYRINGE AND TUBING NOT CONNECTED PROPERLY. THE FSE CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY DEPRESSED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A POSITIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED TROPONIN I RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION CLINICAL CHEMISTRY SYSTEM CLINICAL CHEMISTRY SYSTEM JJE SIEMENS HEALTHCARE DIAGNOSTICS, INC. RXLMAXRH NA

Patients

Seq Age Sex Outcome Treatment