FDA Adverse Event Malfunction Summary report: N

AUTOFEED VENTED HUMIDIFICTION CHAMBER

MDR report key: 1172321 · Received September 19, 2008

Report

Report Number
9611451-2008-00541
Event Type
Malfunction
Date Received
September 19, 2008
Date of Event
August 12, 2008
Report Date
August 19, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED CHAMBER (1 UNIT) WAS INVERTED AND BOTH PRIMARY AND SECONDARY FLOATS OBSERVED FOR MOVEMENT. RESULTS: BOTH FLOATS REMAINED FIXED TO THE ALUMINIUM BASE WHEN THEY SHOULD HAVE MOVED. UPON INSPECTION, A LARGE BOW WAS OBSERVED ON THE EDGE OF THE ALUMINIUM BASE IMMEDIATELY TO THE LEFT OF THE HINGE BRACKET. THERE IS ALSO CRAZING IN THIS AREA. CONCLUSION: CHAMBER OVERFILLING IS A KNOWN FAILURE MODE. IMPACT DAMAGE TO THE CHAMBER, FOR INSTANCE BY DROPPING, CAN LEAD TO MISALIGNMENT AND SUBSEQUENT JAMMING OF THE VALVE FLOAT MECHANISM. THIS PREVENTS FLOAT OPERATION, ALLOWING WATER TO FILL UNCONTROLLED IN THE CHAMBER. THE BOW OBSERVED IN THE ALUMINIUM BASE INDICATES AN IMPACT. OUR INSTRUCTIONS FOR USE INDICATE IN PICTORIAL FORM THE MAXIMUM FILL LINE AND ADVISES THE USER TO REPLACE THE CHAMBER SHOULD WATER EXCEED THE LIMIT LINE. FISHER & PAYKEL HEALTHCARE HAS AN OPEN CAPA ITEM TO ADDRESS MR290 CHAMBER OVERFILLING. OUR MONITORING AND TRENDING OF CHAMBER OVERFILLING HAS A RATE OF OCCURRENCE WORLDWIDE OF 0.0008% IN THE LAST YEAR TO AUGUST 2008.

Description of Event or Problem · 1

A HOSPITAL REPORTED TO OUR DISTRIBUTOR THAT 2 UNITS OF THE MR290V AUTOFEED VENTED HUMIDIFICATION CHAMBER CONNECTED TO AN RT235 BREATHING CIRCUIT, OVERFILLED WITH WATER AS BOTH THE FLOATS IN THE CHAMBERS DID NOT WORK. WATER SUBSEQUENTLY ENTERED THE BREATHING CIRCUIT. ONE OF THE AFFECTED CHAMBERS WAS DISCARDED. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOFEED VENTED HUMIDIFICTION CHAMBER BTT FISHER & PAYKEL HEALTHCARE, LTD. MR290V 080418

Patients

Seq Age Sex Outcome Treatment
1 INFANT EVAQUA BREATHING CIRCUIT