INFANT EVAQUA BREATHING CIRCUIT

MDR report key: 1172312 · Received September 19, 2008

Report

Report Number
9611451-2008-00544
Event Type
Malfunction
Date Received
September 19, 2008
Report Date
August 25, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BZE
PMA / PMN Number
K034026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B) (4). METHOD: ONLY 1 OF THE 3 RT235 BREATHING CIRCUITS WAS RETURNED TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. THE RETURNED DEVICE WAS NOT A WHOLE CIRCUIT BUT HAD BEEN CUT OFF AT ONE END. AN ATTEMPT WAS MADE TO INSERT A HEATER WIRE ADAPTER TO THE HEATER WIRE SOCKET OF THE INSPIRATORY TUBE. RESULTS: THE HEATED BREATHING CIRCUIT CONTAINS A HEATER WIRE SOCKET FOR CONNECTION TO THE HUMIDIFIER. IN THIS CASE, THE HEATER WIRE ADAPTER COULD NOT BE INSERTED AS ONE OF THE HEATER WIRE PINS WAS BENT. UPON FURTHER INSPECTION, THE TIP OF THE PIN WAS DISCOVERED TO BE SLIGHTLY CRUSHED. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR THIS LOT NUMBER. CONCLUSION: FISHER & PAYKEL HEALTHCARE IMPLEMENTED IMPROVEMENTS TO THE PRODUCTION PROCESS IN RESPECT OF INFANT BREATHING CIRCUITS IN MAY 2007 WITH THE AIM OF PREVENTING BENT PINS FROM PASSING THE PRODUCTION TEST. THIS IS ACHIEVED THROUGH PRIOR IDENTIFICATION AND REJECTION OF DAMAGED HEATER WIRE PINS. THIS EVENT OCCURRED ON A DEVICE AFTER THE PRODUCTION IMPROVEMENTS. IT IS POSSIBLE THAT A DAMAGED PIN, WHILST PASSING THE FINAL ELECTRICAL RESISTANCE PRODUCTION TEST, CAN BE FURTHER DAMAGED DURING SET-UP AND CONNECTION OF THE EQUIPMENT WHEN USED. (B) (4).

Description of Event or Problem · 1

OUR DISTRIBUTOR IN (B) (6) REPORTED TO A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT THE ELECTRICAL HEATER WIRE SOCKET IN THE RT235 INFANT EVAQUA BREATHING CIRCUIT COULD NOT BE CONNECTED TO THE HUMIDIFIER. THIS WAS OBSERVED IN 3 DEVICES. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT EVAQUA BREATHING CIRCUIT BZE FISHER & PAYKEL HEALTHCARE, LTD. RT235 080514

Patients

Seq Age Sex Outcome Treatment
1