SEFANT CONTINOUS FLOW BREATHING CIRCUIT
Report
- Report Number
- 9611451-2008-00546
- Event Type
- Malfunction
- Date Received
- September 19, 2008
- Report Date
- August 27, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- BTT
- PMA / PMN Number
- K020332
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
NO COMPLAINT DEVICES WERE RETURNED TO FISHER & PAYKEL HEALTHCARE ("FPH") FOR INVESTIGATION. AN FPH CLINICAL PRODUCT SPECIALIST MADE A SITE VISIT TO THE NEONATAL INTENSIVE CARE UNIT OF THE HOSPITAL IN 2008. THE PRODUCT SPECIALIST OBSERVED A TOTAL OF 6 EQUIPMENT SET-UPS ON BOTH VENTILATOR AND SIPAP APPLICATIONS. RESULTS: NONE OF THE 6 EQUIPMENT SET-UPS WERE CORRECT. NO TEMPERATURE PROBE COVERS (REFLECTIVE SHIELDS) WERE UTILIZED IN THE SET-UPS INVOLVING INFANT RADIANT WARMERS. IN ADDITION, TEMPERATURE PROBE COVERS AND UNHEATED EXTENSIONS WERE NOT PROPERLY USED, AND THE PLACEMENT OF THE AIRWAY TEMPERATURE PROBES WAS INCORRECT IN SET-UPS INVOLVING INCUBATORS. CONCLUSION: THE PURPOSE OF THE REFLECTIVE SHIELD IS TO PROTECT THE AIRWAY TEMPERATURE PROBE FROM RADIANT HEAT FROM THE INFANT WARMER. WITHOUT THIS PROTECTION, HEAT FROM THE INFANT WARMER CAN CAUSE THE TEMPERATURE PROBE TO GIVE AN ERRONEOUSLY HIGH READING, WHICH SIGNALS THE MR850 HUMIDIFIER TO PRODUCE LESS HEAT. THE REDUCTION IN HEATING WITHIN THE BREATHING CIRCUIT CAN CAUSE AN INCREASE IN THE AMOUNT OF CONDENSATION FORMING. FPH'S PRODUCT SPECIALIST RE-EDUCATED HOSPITAL STAFF ON THE CORRECT USE OF REFLECTIVE SHIELDS, UNHEATED EXTENSIONS AND CORRECT TEMPERATURE PROBE PLACEMENT. THE FPH RT020 OR MAQUET EXPIRATORY FILTER AND THE RT130 BREATHING CIRCUIT WHICH CONTAINS A WATER TRAP WERE RECOMMENDED. THE HOSPITAL, HAVING ADOPTED THE CHANGES, SUBSEQUENTLY REPORTED FAVORABLE RESULTS.
REPORTER REPORTED TO A FISHER & PAYKEL HEALTHCARE CLINICAL PRODUCT SPECIALIST OF EXCESSIVE RAINOUT (CONDENSATION) OCCURRING IN THE NEONATAL INTENSIVE CARE UNIT WITH SET-UPS CONSISTING OF THE RT132 INFANT CONTINUOUS FLOW BREATHING CIRCUIT IN APPLICATIONS UTILISING SIPAP, CARDINAL BREATHING CIRCUIT WITH, AND WITHOUT THE AEROGEN SOLO NEBULIZER, MR850 RESPIRATORY HUMIDIFIER AND THE SERVO-I VENTILATOR. THE HOSPITAL FURTHER REPORTED THE FOLLOWING: THAT EXPIRATORY RAINOUT WAS CAUSING THE EXPIRATORY FILTER AND EXPIRATORY BLOCK/CARTRIDGE TO BECOME WET. THAT STAFF HAD BEEN ADVISED BY ANOTHER DISTRIBUTOR TO MOUNT THE AEROGEN NEBULIZER AT THE HUMIDIFICATION CHAMBER, NOT IN THE AREA BETWEEN THE AIRWAY TEMPERATURE PROBE AND BREATHING CIRCUIT WYE AS RECOMMENDED BY FISHER & PAYKEL HEALTHCARE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SEFANT CONTINOUS FLOW BREATHING CIRCUIT | BTT | FISHER & PAYKEL HEALTHCARE, LTD. | RT132 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARDINAL BREATHING CIRCUIT| MR850 HUMIDIFIER| AEROGEN SOLO NEBULIZER| SERVI-I VENTILATOR |