PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2008-01287
- Event Type
- Malfunction
- Date Received
- September 19, 2008
- Report Date
- September 18, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- UNKNOWN
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING DISCOVERED A BROKEN PITCH CABLE AND A BROKEN WIRE AT THE YAW PULLEY EXIT, WHERE A PIECE OF MATERIAL IS ALSO MISSING. ENGINEERING ALSO OBSERVED DEEP SCRATCHES ON A LARGE AREA AROUND THE MAIN TUBE, WHERE MATERIAL HAS BEEN REMOVED. THE SCRATCHES EXTEND APPROXIMATELY 1.750" FROM THE INTERSECTION WITH THE PROXIMAL CLEVIS. BASED ON THE LOCATION AND APPEARANCE, THE DAMAGE WAS MOST LIKELY CAUSED BY INSTRUMENT COLLISIONS. NO OTHER DAMAGE FOUND. DUE TO THE DAMAGE OBSERVED ON THE MAIN TUBE, IT'S POSSIBLE THAT THE WIRE AND PITCH CABLE BREAKAGES MAY HAVE ALSO BEEN A RESULT OF INSTRUMENT COLLISIONS. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT DURING A DA VINCI S TOTAL HYSTERECTOMY SURGICAL PROCEDURE, THE JAWS OF THE PK DISSECTING FORCEPS INSTRUMENT WOULD NOT CLOSE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED, AND THE PLANNED SURGICAL PROCEDURE WAS COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PK DISSECTING FORCEPS INSTRUMENT | ELECTROSURGICAL INSTRUMENT | GEI | INTUITIVE SURGICAL, INC. | 420227-02 | S10080707 195 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | AN ELECTROSURGICAL UNIT| ACCESSORIES| DA VINCI S SURGICAL SYSTEM |