ANALYTICAL P MODULE
MDR report key: 1172269
·
Received September 19, 2008
Report
- Report Number
- 1823260-2008-07025
- Event Type
- Malfunction
- Date Received
- September 19, 2008
- Date of Event
- September 2, 2008
- Report Date
- September 19, 2008
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K953239
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Description of Event or Problem · 1
USER RECEIVED DISCREPANT PHOSPHORUS RESULTS FOR ONE PATIENT SAMPLE. INITIAL RESULT 17.9 MG/DL. USER STATES THEIR LIS SYSTEM CAUGHT THIS HIGH RESULT AND THEY DECIDED TO REPEAT THE SAMPLE. REPEAT RESULT 4.6 MG/DL. INITIAL RESULT WAS NOT REPORTED. THE FIELD SERVICE REP DETERMINED THERE WAS A BAD R2 PROBE AND REPLACED AND ADJUSTED IT. PERFORMANCE TESTS WERE PERFORMED WHICH WERE WITHIN SPECIFICATION.