FDA Adverse Event Injury Summary report: N

DYONICS PWR HIGH SPEED DRILL

MDR report key: 11722134 · Received April 25, 2021

Report

Report Number
1643264-2021-01533
Event Type
Injury
Date Received
April 25, 2021
Date of Event
March 24, 2021
Report Date
July 9, 2021
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HAB
UDI-DI
03596010466167
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE: CASE-(B)(4). H6: THE REPORTED DEVICE, USED IN TREATMENT, WAS RECEIVED FOR EVALUATION. THERE WAS A RELATIONSHIP FOUND BETWEEN THE RETURNED DEVICE AND THE REPORTED INCIDENT. A VISUAL INSPECTION WAS PERFORMED AND NO DEFICIENCIES WERE OBSERVED. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME. A FUNCTIONAL EVALUATION REVEALED A BLADE STALL ERROR MESSAGE AND EXCESSIVE HEAT PRODUCED FROM THE HANDPIECE. AN INVESTIGATION PERFORMED BY THE SUPPLIER REVEALED THE DEVICE FAILED THE TORQUE TEST AND HAS A GRINDING NOISE. THE COMPLAINT WAS CONFIRMED AND THE ROOT CAUSE WAS ASSOCIATED WITH A MECHANICAL COMPONENT FAILURE. FACTORS WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT, INCLUDE CORROSION IN THE GEARBOX/MOTOR ASSEMBLY FROM CLEANING AND STERILIZATION METHODS AND THE CHEMICALS INVOLVED OVER A PERIOD OF TIME. A BLADE STALL CONDITION CAN RESULT IN INCREASED CURRENT DRAW FROM THE CONTROL UNIT WHICH WILL HEAT THE MOTOR AND HAND PIECE HOUSING. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW CONCLUDED THIS WAS A REPEAT ISSUE. THE INSTRUCTIONS FOR USE WAS REVIEWED AND FOUND TO INCLUDE CONDITIONS OF OFF LABEL USE AND TECHNIQUE SPECIFICS, AS WELL AS PRECAUTIONS AND WARNINGS RELATED TO THE USE OF THE DEVICE. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME.

Additional Manufacturer Narrative · 1

INTERNAL COMPLAINT REFERENCE (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IP JOINT FUSION THE DYONICS PWR HIGH SPEED DRILL DID NOT HAVE ANY POWER. THIS EVENT HAPPENED WITH A DYONICS PWR HIGH SPEED DRILL. HAND PIECE CONTROL ERROR MESSAGE WAS DISPLAYED IN THE SCREEN. INSTRUMENTS INSIDE PATIENT. SURGICAL DELAY OF LESS THAN OR EQUAL TO 30 MINUTES. PROCEDURE FINISHED WITH COMPETITOR DEVICE. NO PATIENT COMPLICATIONS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
618202 DYONICS PWR HIGH SPEED DRILL SAW, POWERED, AND ACCESSORIES HAB SMITH & NEPHEW, INC. 03596010466167

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention