FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 11721480 · Received April 23, 2021

Report

Report Number
1221359-2021-01235
Event Type
Malfunction
Date Received
April 23, 2021
Report Date
June 17, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCABOROUGH
Product Code
QJR
UDI-DI
10811877011269
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: INVESTIGATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4) INC. ON RETAINED KIT LOT 1023084 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: 1023084 , TEST BASE PART NUMBER 190-000 / LOT: 1023084. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. THE CURRENT OVERALL INCIDENT RATE FOR FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) FOR THIS SPECIFIC LOT BASED ON THE TOTAL QUANTITY OF DEVICES MANUFACTURED FOR DISTRIBUTION IS (B)(4). REPLACE THIS LAST SENTENCE WITH: ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED DURING ROUTINE PATIENT TESTING, FIVE FALSE POSITIVE RESULT WITH ID NOW COVID 19 ASSAY USING A DIRECT NASOPHARYNGEAL SWAB. REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING ON NASOPHARYNGEAL SWABS CT VALUES NOT PROVIDED. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
616567 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19, QJR ABBOTT DIAGNOSTICS SCABOROUGH 1023084 10811877011269

Patients

Seq Age Sex Outcome Treatment
1