FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 11721471 · Received April 23, 2021

Report

Report Number
1221359-2021-01249
Event Type
Malfunction
Date Received
April 23, 2021
Date of Event
March 24, 2021
Report Date
January 27, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE INVESTIGATION CONCLUSION. PLEASE SEE UPDATES: G3, G6, H2 AND H6. ADDITIONAL INFORMATION: H10. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1009983 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT 1009983 AND TEST BASE PART NUMBER 190-430 / LOT 1009983.THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1009983 SHOWED THAT THE COMPLAINT RATE IS 0.11%. IN CONCLUSION, THE MANUFACTURING BATCH RECORD REVIEW REVEALED THAT THE PRODUCT MET ACCEPTANCE CRITERIA FOR RELEASE. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE DELETED PRIOR TO EXPORT. REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES PRODUCT PERFORMANCE IS AS EXPECTED AND HAVE NOT IDENTIFIED A PRODUCT DEFICIENCY. BASED ON THE ABOVE SUMMARY, THE INVESTIGATION IS DEEMED CLOSED. THE PRODUCT WILL CONTINUE TO BE MONITORED AND TRACKED.

Additional Manufacturer Narrative · 1

REFERENCE REPORTS: 1221359-2021-01243, 1221359-2021-01244, 1221359-2021-01245, 1221359-2021-01246, 1221359-2021-01247, 1221359-2021-01248. THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY FOR MULTIPLE PATIENTS PERFORMED ON VARIOUS DATES. THIS REPORT ADDRESSES EVENT SEVEN OF SEVEN . THE CUSTOMER REPORTED EIGHTEEN FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A DIRECT TESTED THROAT AND NASAL PURITAN "STERILE FOAM TIPPED APPLICATOR" REF 25-1506 1PF 100 SWABS. IT IS UNKNOWN IF REPEAT TESTING WAS PERFORMED. CONFIRMATION TESTING WAS PERFORMED ON NASAL AND THROAT SAMPLES WITH PCR GENERATED NEGATIVE RESULTS; CT VALUES NOT PROVIDED. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
616068 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1009983 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 Unknown