FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE VOLUMETRIC INFUSION PUMP
MDR report key: 1172143
·
Received September 25, 2008
Report
- Report Number
- 6000001-2007-02851
- Event Type
- Malfunction
- Date Received
- September 25, 2008
- Date of Event
- January 4, 2007
- Report Date
- January 4, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 25, 2007. EVALUATION SUMMARY:FALIURE CODE 812:02 WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THIS WAS CAUSED BY AN OUTDATED SHUTTLE MOTOR THAT WAS REPLACED.
Description of Event or Problem · 1
DURING PRODUCT EVALUATION, FAILURE CODE 812:02 WAS FOUND IN THE PUMP'S EVENT HISTORY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |