HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER
Report
- Report Number
- 1226348-2008-00247
- Event Type
- Injury
- Date Received
- September 23, 2008
- Date of Event
- August 25, 2008
- Manufacturer
- CODMAN & SHURTLEFF, INC./MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- K974739
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICES WHICH HAVE INTERFERED WITH THEIR PROGRAMMING AND/OR PRESSURE CONTROL MECHANISMS. THE BIOLOGICAL DEBRIS HAS CAUSED THE RETURNED VALVES TO FAIL THE PROGRAMMING AND/OR PRESSURE TESTS AND APPEARS TO HAVE CAUSED THE DIFFICULTIES EXPERIENCED BY THE CUSTOMERS. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFO REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.
AFFILIATE REPORTED THAT THE DEVICE WOULD NOT REPROGRAM AFTER IMPLANTATION LEAVING THE SURGEON DOUBTFUL OF THE ACCURACY OF THE PRESSURE SETTING AS WELL AS THE AMOUNT OF FLUID FLOWING THROUGH THE VALVE. THE DEVICE NEEDED TO BE REPLACED. THE PROGRAMMER 82-3120 WAS USED. THE DEVICE WAS IMPLANTED IN 2001, AND REPLACED IN 2008.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC./MEDOS S.A. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR | Required Intervention |