AFFINITY BED

MDR report key: 1171974 · Received September 18, 2008

Report

Report Number
1824206-2008-03510
Event Type
Malfunction
Date Received
September 18, 2008
Date of Event
August 20, 2008
Report Date
August 20, 2008
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN WAS PERFORMING PREVENTIVE MAINTENANCE ON THE BIRTHING BED AND FOUND THE HEAD SECTION WOULD DRIFT. THE TECHNICIAN CLEANED THE HEAD DRIVE BRAKE AND REPLACED THE BROKEN SPRING TO REPAIR THE BED. CHAPTER 6 OF THE SERVICE MANUAL DOCUMENTS A FUNCTION CHECK FOR HEAD DRIVE SCREW AS PART OF PREVENTATIVE MAINTANANCE.

Description of Event or Problem · 1

HEAD SECTION FUNCTION OF THE BED IS DRIFTING DOWN.