FDA Adverse Event
Malfunction
Summary report: N
TRANSTAR STRETCHER
MDR report key: 1171943
·
Received September 19, 2008
Report
- Report Number
- 1824206-2008-03531
- Event Type
- Malfunction
- Date Received
- September 19, 2008
- Date of Event
- August 28, 2008
- Report Date
- August 28, 2008
- Manufacturer
- HILL-ROM
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE BRAKES NOT HOLDING/WORKING COULD LEAD TO UNINTENTIONAL MOVEMENT OF THE STRETCHER, WHICH HAS THE POTENTIAL TO CAUSE SERIOUS INJURY. THE TECHNICIAN REPLACED THE CASTERS IN ORDER TO RESOLVE THE PROBLEM.
Description of Event or Problem · 1
A TECHNICIAN FOUND THAT THE BRAKES WOULD NO LONGER HOLD ON THIS STRETCHER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSTAR STRETCHER | WHEELED STRETCHER | FPO | HILL-ROM | P8020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |