FDA Adverse Event Malfunction Summary report: N

LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 1171935 · Received September 19, 2008

Report

Report Number
3015876-2008-01230
Event Type
Malfunction
Date Received
September 19, 2008
Date of Event
August 22, 2008
Report Date
August 22, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K991910
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE SYSTEM/MEMORY PCB ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO-CONTROL FURTHER EVALUATED THE REPLACED ASSEMBLY AND OBSERVED THAT A CONNECTOR, DESIGNATOR J3, HAD A LOOSE CONNECTION, CAUSING THE DEVICE TO RESET.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE POWER IS RESETTING., THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA