FDA Adverse Event Malfunction Summary report: N

VERSACARE

MDR report key: 1171928 · Received September 19, 2008

Report

Report Number
1824206-2008-03497
Event Type
Malfunction
Date Received
September 19, 2008
Date of Event
August 22, 2008
Report Date
August 22, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RECEIVED EMAIL FROM HILL-ROM REGIONAL SALES MANAGER STATING THAT HE HAS BEEN INFORMED THAT THE TWO NEW LOAD BEAMS, AND THE NEW SCALE PC BOARD HAVE BEEN INSTALLED AND THE SCALE IS NOW FUNCTIONING PROPERLY. THIS RESOLVED THE ISSUE OF THE SCALE NOT WEIGHING PROPERLY.

Description of Event or Problem · 1

RECEIVED EMAIL FROM CUSTOMER STATING THAT THE SCALE IS NOT WEIGHING PROPERLY. HE STATED THAT THE WEIGHT IS RANDOMLY OFF 3 TO 7 KG UP OR DOWN EVERY TIME. HE CONFIRMED THAT THE PACKING SHIMS HAVE BEEN REMOVED. HE CLAIMED THAT P1 RED LED IS LIT ON THE SCALE PC BOARD AND HE SWAPPED P1 AND P2 CONNECTORS AND P1 REMAINED LIT. HE ALSO STATED THAT THE SCALE IS DISPLAYING ERR3. I ASKED HIM TO CHECK THE WEIGHT FRAME TO SEE IF IT IS BENT. HE CLAIMS THAT IT IS NOT BENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSACARE AC POWERED HOSPITAL BED FNL HILL-ROM RITTER P3200 NA

Patients

Seq Age Sex Outcome Treatment
1