FDA Adverse Event Malfunction Summary report: N

CENTRISOL HEMODIALYSIS ACID CONCENTRATE

MDR report key: 11718016 · Received April 23, 2021

Report

Report Number
2150060-2021-00006
Event Type
Malfunction
Date Received
April 23, 2021
Date of Event
April 9, 2021
Report Date
April 23, 2021
Manufacturer
MEDIVATORS
Product Code
FKQ
UDI-DI
00677964000577
PMA / PMN Number
K843963
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE USER FACILITY REPORTED POSITIVE DETECTION OF OXIDIZING AGENTS IN THEIR DIALYSATE USING CENTRISOL ACID CONCENTRATE LOT 489224.FURTHER CUSTOMER TESTING CONFIRMED THE PRESENCE OF OXIDIZING AGENTS IN THE CENTRISOL ACID CONCENTRATE. INVESTIGATIONS PERFORMED BY MEDIVATORS ON RETURNED PRODUCTS CONFIRMED THAT A LIMITED SUBSET OF BOTTLES FROM THE IMPACTED LOT CONTAIN AN OUT-OF-SPECIFICATION CONCENTRATION FORMULATION DUE TO THE PRESENCE OF AN OXIDIZING AGENT.. A RECALL OF ALL PRODUCTS FROM THE IMPACTED LOT WILL BE CONDUCTED AND THE SITUATION WILL BE COMMUNICATED TO IMPACTED CUSTOMERS. THE USE OF OUT-OF-SPECIFICATION FORMULATIONS OF THE CONCENTRATE MAY PRESENT A RISK WHEN USED IN PATIENT PROCEDURES. HOWEVER, BOTH TEST STRIP USAGE AND CONDUCTIVITY CHECKS ARE INDICATED AS STANDARD CLINICAL PRACTICE AND WOULD MAKE THE SITUATION DETECTABLE TO THE USER. NO HARM HAS BEEN REPORTED AS A RESULT OF THIS SITUATION. MEDIVATORS HAS OPENED CAPA-(B)(4). TO ADDRESS THIS SITUATION. MEDIVATORS WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS TO ENSURE THE PRODUCT PERFORMS AS EXPECTED.

Description of Event or Problem · 1

THE USER FACILITY REPORTED POSITIVE DETECTION OF OXIDIZING AGENTS IN THEIR DIALYSATE USING CENTRISOL ACID CONCENTRATE LOT 489224. FURTHER CUSTOMER TESTING CONFIRMED THE PRESENCE OF OXIDIZING AGENTS IN THE CENTRISOL ACID CONCENTRATE. NO HARM HAS BEEN REPORTED AS A RESULT OF THIS SITUATION.