FDA Adverse Event Injury Summary report: N

ZEPHYR ENDOBRONCHIAL VALVE (EBV)

MDR report key: 11717114 · Received April 23, 2021

Report

Report Number
3007797756-2021-00071
Event Type
Injury
Date Received
April 23, 2021
Date of Event
March 16, 2021
Report Date
July 27, 2021
Product Code
NJK
UDI-DI
00811907030423
PMA / PMN Number
P180002
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

HEATH EFFECT CLINICAL CODE 4581 CHOSEN FOR SUBCUTANEOUS EMPHYSEMA.

Description of Event or Problem · 0

SUBJECT HAD ZEPHYR VALVES IMPLANTED ON (B)(6) 2021. SUBJECT EXHIBITED A SMALL, SYMPTOMATIC PNEUMOTHORAX AND A CHEST TUBE WAS PLACED ON (B)(6) 2021. A VALVE WAS REMOVED ON (B)(6)2021. THE CHEST TUBE WAS REMOVED. ON (B)(6) 2021 THE SUBJECT HAD A MODERATE, SYMPTOMATIC, LEFT PNEUMOTHORAX AND A CHEST TUBE WAS PLACED. THE SUBJECT THEN EXPERIENCED SUBCUTANEOUS EMPHYSEMA SEEN SPREADING TO THE FACE. THE SUBJECT WAS TREATED WITH A LEFT ANTERIOR BLOW HOLE PROCEDURE. THE CHEST TUBE WAS REMOVED AND THE SUBJECT WAS DISCHARGED ON (B)(6) 2021.

Additional Manufacturer Narrative · 1

PNEUMOTHORAX IS THE MOST COMMON SIDE EFFECT ASSOCIATED WITH THE ZEPHYR VALVE TREATMENT (CRINER ET AL. A MULTICENTER RANDOMIZED CONTROLLED TRIAL OF ZEPHYR ENDOBRONCHIAL VALVE TREATMENT IN HETEROGENEOUS EMPHYSEMA (LIBERATE). AM J RESPIR CRIT CARE MED. 2018; 198 (9): 1151-1164). TARGETED LOBAR DEFLATION LIKELY CAUSES INFLATION OF THE IPSILATERAL LOBE, WHICH CAN RESULT IN A TEAR OF THE ALREADY COMPROMISED PARENCHYMAL TISSUE OF THE EMPHYSEMATOUS IPSILATERAL LOBE, RESULTING IN A PNEUMOTHORAX (CRINER ET AL. A MULTICENTER RANDOMIZED CONTROLLED TRIAL OF ZEPHYR ENDOBRONCHIAL VALVE TREATMENT IN HETEROGENEOUS EMPHYSEMA (LIBERATE). AM J RESPIR CRIT CARE MED. 2018; 198 (9): 1151-1164). IN THE LIBERATE STUDY (IDE CLINICAL STUDY USED TO SUPPORT PMA P180002'S APPROVAL), 26.6% OF THE ZEPHYR VALVE SUBJECTS EXPERIENCED A PNEUMOTHORAX IN THE TREATMENT PERIOD ([LESS THAN OR EQUAL TO 45 DAYS). THESE WERE MANAGED USING STANDARD OF CARE PROCEDURES AS PER PREVIOUSLY PUBLISHED GUIDELINES (VALIPOUR, ARSCHANG, ET AL. RESPIRATION 87.6 (2014): 513-521). IN 17.4% OF THE EVENTS, THE PNEUMOTHORAX RESOLVED WITHOUT ANY ADDITIONAL INTERVENTION WITH SUBJECTS UNDER CAREFUL OBSERVATION. IN OVER HALF THE EVENTS (56.5%), THE PNEUMOTHORAX WAS MANAGED WITH A CHEST-TUBE ONLY. AN ADDITIONAL 13% OF THE EVENTS WERE MANAGED WITH A CHEST-TUBE AND THE TEMPORARY REMOVAL OF ONE OR MORE VALVES, WHILE ANOTHER 13% OF THE EVENTS WERE MANAGED WITH A CHEST-TUBE AND REMOVAL OF ALL THE IMPLANTED VALVES. UPON SUCCESSFUL RESOLUTION OF THE PNEUMOTHORAX, REMOVED VALVES CAN BE REPLACED. PATIENTS THAT EXPERIENCED A PNEUMOTHORAX EXPERIENCED CLINICAL BENEFITS OF THE ZEPHYR VALVE TREATMENT THAT WERE SIMILAR TO THE BENEFITS EXPERIENCED BY PATIENTS WHO DID NOT HAVE A PNEUMOTHORAX. THE ZEPHYR EBV SYSTEM IFU AND PULMONX TRAINING PROGRAM BOTH SPECIFICALLY REFERENCE PNEUMOTHORAX AS A KNOWN SIDE EFFECT OF THIS PROCEDURE AND THE PUBLISHED GUIDELINES (VALIPOUR, ARSCHANG, ET AL. "EXPERT STATEMENT: PNEUMOTHORAX ASSOCIATED WITH ENDOSCOPIC VALVE THERAPY FOR EMPHYSEMA-POTENTIAL MECHANISMS, TREATMENT ALGORITHM, AND CASE EXAMPLES." RESPIRATION 87.6 (2014): 513-521). THE REPORTED EVENT ALIGNS WITH THE EXPERIENCE OBSERVED IN THE LIBERATE CLINICAL STUDY AND IS AN EXPECTED SIDE EFFECT TO THE ZEPHYR VALVE TREATMENT.

Description of Event or Problem · 1

SUBJECT HAD ZEPHYR VALVES IMPLANTED ON (B)(6) 2021. SUBJECT EXHIBITED A SMALL, SYMPTOMATIC PNEUMOTHORAX AND A CHEST TUBE WAS PLACED ON (B)(6) 2021. A VALVE WAS REMOVED ON (B)(6) 2021. THE CHEST TUBE WAS REMOVED ON (B)(6) 2021. THE SUBJECT WAS DISCHARGED ON (B)(6) 2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
608938 ZEPHYR ENDOBRONCHIAL VALVE (EBV) ENDOBRONCHIAL VALVE NJK ZEPHYR 5.5-LP EBV 504831-V7.0 00811907030423

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R