FDA Adverse Event Malfunction Summary report: N

BD FACSLYRIC

MDR report key: 11716603 · Received April 23, 2021

Report

Report Number
2916837-2021-00193
Event Type
Malfunction
Date Received
April 23, 2021
Date of Event
April 4, 2021
Report Date
May 5, 2021
Manufacturer
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Product Code
OYE
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION OF THE COMPLAINT, IT HAS BEEN DETERMINED THAT THE PREVIOUSLY SUBMITTED MFR REPORT# 2916837-2021-00193 WAS SENT IN ERROR. THE FSE CONFIRMED THAT THE LEAKAGE WAS CONTAINED WITHIN INSTRUMENT, THEREFORE, THIS IS NOT CONSIDERED TO BE A REPORTABLE MALFUNCTION.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD FACSLYRIC¿ WASTE LEAKED OUTSIDE OF INSTRUMENT. THERE WAS NO USER IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THERE ARE PQC FAILURES AND FLUID IS DRIPPED FROM THE SAMPLE PROBE DURING CLEANING. WAS THE LEAK CONTAINED WITHIN THE INSTRUMENT? NO. WAS THE LEAK IN A CUSTOMER ACCESSIBLE LOCATION? YES. WHAT WAS THE FLUID THAT LEAKED? WASTE. WHAT IS THE SOURCE OF LEAK -- WASTE LINE OR NON-WASTE LINE? WASTE LINE. WAS THE CUSTOMER EXPOSED TO BLOOD OR BODILY FLUIDS? NO. WAS THERE ANY PHYSICAL HARM TO THE CUSTOMER AS A RESULT OF THE LEAK? NO. WAS THE FLUID SPRAYING OUT FROM THE POINT OF THE LEAK? NO. WAS THE WASTE MIXED WITH DECONTAMINATION / BLEACH SOLUTION? UNKNOWN. WAS PROPER PPE BEING WORN AT TIME OF LEAK AND DURING CLEAN UP? YES.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MEDICAL DEVICE EXPIRATION DATE: NA. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING BD FACSLYRIC¿ WASTE LEAKED OUTSIDE OF INSTRUMENT. THERE WAS NO USER IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THERE ARE PQC FAILURES AND FLUID IS DRIPPED FROM THE SAMPLE PROBE DURING CLEANING. WAS THE LEAK CONTAINED WITHIN THE INSTRUMENT? NO. WAS THE LEAK IN A CUSTOMER ACCESSIBLE LOCATION? YES. WHAT WAS THE FLUID THAT LEAKED? WASTE. WHAT IS THE SOURCE OF LEAK -- WASTE LINE OR NON-WASTE LINE? WASTE LINE. WAS THE CUSTOMER EXPOSED TO BLOOD OR BODILY FLUIDS? NO. WAS THERE ANY PHYSICAL HARM TO THE CUSTOMER AS A RESULT OF THE LEAK? NO. WAS THE FLUID SPRAYING OUT FROM THE POINT OF THE LEAK? NO. WAS THE WASTE MIXED WITH DECONTAMINATION / BLEACH SOLUTION? UNKNOWN. WAS PROPER PPE BEING WORN AT TIME OF LEAK AND DURING CLEAN UP? YES.