IMP,TSV,4.1MM,SBM,11.5
Report
- Report Number
- 0002023141-2021-01033
- Event Type
- Injury
- Date Received
- April 23, 2021
- Date of Event
- April 5, 2021
- Report Date
- July 23, 2021
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019232
- PMA / PMN Number
- K072589
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- DENTIST
Narratives
SIMILAR COMPLAINTS FOR THE INABILITY TO ACHIEVE PRIMARY STABILITY (UNABLE TO PLACE IMPLANT INTO OSTEOTOMY) HAVE BEEN PREVIOUSLY INVESTIGATED. REFER TO ATTACHED SUMMARY INVESTIGATION. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORDS REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH HAVE NOT SUGGESTED THE LIKELY RELEASE OF NON-CONFORMING PRODUCT. TO DATE, ALL COMPLAINT DATA WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/IE ESCALATION. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBERS (63706006 AND 1222559). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBERS (63706006 AND 1222559) FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: UNABLE TO PLACE IMPLANT INTO OSTEOTOMY-. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THIS REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: G6: CHECKED "FOLLOW-UP". H2: CHECKED FOLLOW-UP TYPE. H3: CHANGED "NO" TO "YES".
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT WEIGHT UNKNOWN / NOT PROVIDED. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR LACK OF PRIMARY STABILITY EVENTS THAT DO NOT ALLEGE A DEFICIENCY WITH THE IMPLANT AND IDENTIFIED THAT THE REPORTED EVENT IS LIKELY DUE TO BIOLOGICAL FACTORS WHICH HAVE AN ADVERSE EFFECT ON IMPLANT STABILITY OR IS RELATED TO SURGICAL TECHNIQUE AND INSERTION TORQUE. DUE TO A WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED), IDENTIFYING A DEFINITIVE ROOT CAUSE IS GENERALLY NOT POSSIBLE. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. SUMMARY REPORTING.
IT WAS REPORTED THAT IMPLANT WAS UNABLE TO BE PLACED AT THE SAME SITE AND WERE UNABLE TO OBTAIN PRIMARY STABILITY AT TOOTH # 20. SITE WAS GRAFTED AND PATIENT WILL RETURN FOR AN IMPLANT AT A FUTURE APPOINTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 610684 | IMP,TSV,4.1MM,SBM,11.5 | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSV4B11 | 1222559 | 00889024019232 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention |