FDA Adverse Event Injury Summary report: N

OSS POLY LOCK PIN

MDR report key: 11716206 · Received April 23, 2021

Report

Report Number
0001825034-2021-01219
Event Type
Injury
Date Received
April 23, 2021
Date of Event
March 12, 2019
Report Date
June 3, 2021
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JDI
UDI-DI
00880304010949
PMA / PMN Number
K002757
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; THEREFORE, VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. MEDICAL RECORDS WERE NOT PROVIDED. DEVICE HISTORY RECORD (DHR) WAS REVIEWED FOR DEVIATIONS AND/ OR ANOMALIES WITH NO RELATED DEVIATIONS / ANOMALIES IDENTIFIED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORT: 0001825034-2021-01217, 0001825034-2021-01218, 0001825034-2021-01220, 0001825034-2021-01221, 0001825034-2021-01222, 0001825034-2021-01223, 0001825034-2021-01224, 0001825034-2021-01225, 0001825034-2021-01226. CONCOMITANT MEDICAL PRODUCTS: OSS REINFORCED YOKE ITEM# 150493 LOT# 211420 OSS TIBIAL POLY BEARING 20MM ITEM# 150414 LOT# 237060 OSS NON-MOD TIB PLATE SHORT 67 ITEM# 150417 LOT# 203610 OSS 7CM SEGMENTAL FEMORAL RT ITEM# 150354 LOT# 756200 OSS POLY TIBIAL BUSHING ITEM# 150476 LOT# 460720 OSS POLY FEMORAL BUSHINGS ITEM# 150477 LOT# 370100 OSS AXLE ITEM# 150480 LOT# 543370 OSS REINFORCED YOKE ITEM# 150493 LOT# 658130 OSS TIBIAL POLY BEARING 22MM ITEM# 150415 LOT# 382120 OSS 3CM ELLIP DIAPHYSEAL SEG ITEM# 150461 LOT# 150610 OSS CEMENTED IM STEM 15MMX90MM ITEM# 150363 LOT# 810360 CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY TWO MONTHS¿ POST IMPLANTATION DUE TO WOUND HEALING PROBLEMS THAT DISENGAGED HIS HINGED FEMUR FROM HIS RIGHT TIBIAL COMPONENT. THIS WAS A RESULT OF HAVING HAD A DISTAL FEMORAL REPLACEMENT WITH ABSENCE OF COLLATERAL LIGAMENTS CAUSING THE FEMUR TO DISENGAGE FROM THE TIBIA WHEN THE KNEE BECAME FLEXED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
616768 OSS POLY LOCK PIN PROSTHESIS, HIP JDI ZIMMER BIOMET, INC. N/A 348850 00880304010949

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R