FDA Adverse Event Injury Summary report: N

ARTHRO BIOINDCTIVE IMPLANT 1 LARGE

MDR report key: 11716077 · Received April 23, 2021

Report

Report Number
3003604053-2021-00168
Event Type
Injury
Date Received
April 23, 2021
Date of Event
April 6, 2021
Report Date
January 16, 2024
Manufacturer
SMITH & NEPHEW, INC.
Product Code
OWY
UDI-DI
00885556732922
PMA / PMN Number
K140300
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

H10: H3, H6: THE REPORTED DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THAT IT IS NOT IN ITS ORIGINAL PACKAGING. THE INSERTION DEVICE AND USED BIOINDUCTIVE IMPLANT WERE RETURNED. THERE IS BIOLOGICAL DEBRIS ON THE RETURNED ITEMS. A FUNCTIONAL ASSESSMENT OF THE INSERTION DEVICE FOUND IT ACTUATES AS INTENDED. THE BIOINDUCTIVE IMPLANT CANNOT BE TESTED AS IT WAS PREVIOUSLY DEPLOYED. A COMPLAINT HISTORY REVIEW FOUND NO SIMILAR REPORTED EVENTS. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION UPON RELEASE FOR DISTRIBUTION. THE INSTRUCTIONS FOR USE WAS REVIEWED AND FOUND TO INCLUDE CONDITIONS OF OFF LABEL USE AND TECHNIQUE SPECIFICS, AS WELL AS PRECAUTIONS AND WARNINGS RELATED TO THE USE OF THE DEVICE. A RISK MANAGEMENT REVIEW FOUND THAT THE REPORTED FAILURE AND/OR HARM WAS DOCUMENTED APPROPRIATELY, AND THERE WERE NO INDICATIONS TO SUGGEST THE ANTICIPATED RISK IS NOT ADEQUATE. THE ROOT CAUSE COULD NOT BE DETERMINED. FACTORS THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT INCLUDE EXCESSIVE FORCE ON THE DEVICE, EXCESSIVE TORQUE ON THE DEVICE, ATTEMPTED CORRECTION OF A DAMAGED DEVICE, OFF-AXIS INSERTION, IMPROPER PREPARATION OF THE INSERTION SITE, OR AN INADVERTENT IMPACT EVENT INCONSISTENT WITH NORMAL USE. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME.

Additional Manufacturer Narrative · 1

INTERNAL COMPLAINT REFERENCE CASE-(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT, DURING A SHOULDER ARTHROSCOPY, AN ARTHRO BIOINDCTIVE IMPLANT LARGE WAS LOADED INTO INSERTER. THE CLAM SHELL WAS A LITTLE STIFF TO REMOVE BUT IMPLANT SEEMED TO BE LOADED PROPERLY. THE DEPLOYMENT OF THE IMPLANT WAS SMOOTH, 5 TENDON ANCHORS WERE PLACED; HOWEVER, THE IMPLANT DID NOT RELEASE FROM THE DELIVERY DEVICE AND PULLED OUT OF THE JOINT. THE PROCEDURE WAS COMPLETED USING A DIFFERENT SIZE BACK-UP DEVICE WITH A DELAY OF 30 MINUTES OR LESS. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611904 ARTHRO BIOINDCTIVE IMPLANT 1 LARGE MESH, SURGICAL, COLLAGEN, ORTHOPAEDICS, REINFORCEMENT OF TENDON OWY SMITH & NEPHEW, INC. RC2AN19J2 00885556732922

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention