FDA Adverse Event Malfunction Summary report: N

4 MOTOR ADVANCE BED

MDR report key: 1171577 · Received September 17, 2008

Report

Report Number
1824206-2008-03447
Event Type
Malfunction
Date Received
September 17, 2008
Date of Event
August 20, 2008
Report Date
August 20, 2008
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE HILL-ROM TECH REPORTED THAT THE BED EXIT ALARM SYSTEM DID NOT FUNCTION ON THIS BED. HE FOUND THAT THE BED EXIT TAPE SWITCHES WERE DEFECTIVE. HE REPLACED THE SWITCHES AND THE BED EXIT SYSTEM FUNCTIONED AS DESIGNED. THERE WERE NO REPORTED ADVERSE EVENTS ASSOCIATED WITH THIS PRODUCT MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 4 MOTOR ADVANCE BED AC POWERED HOSPITAL BED FNL HILL-ROM, INC. 1155 NA

Patients

Seq Age Sex Outcome Treatment
1