FDA Adverse Event
Malfunction
Summary report: N
LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL
MDR report key: 1171565
·
Received September 18, 2008
Report
- Report Number
- 3005075853-2008-01777
- Event Type
- Malfunction
- Date Received
- September 18, 2008
- Date of Event
- August 18, 2008
- Report Date
- August 19, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
JAWS. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE ER420 INSTRUMENT WAS RETURNED WITH THE JAWS OVER TWISTED. THE INSTRUMENT WAS CYCLED AND EJECTED THE REMAINING CLIPS DUE TO THE CONDITION OF THE JAWS. THE INSTRUMENT LOCKED OUT AS INTENDED. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH RECORD WAS REVIEWED, AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS USED DURING A LAPAROSCOPIC CHOLECYSTECTOMY. THE DEVICE JAMMED IN THE NOSE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.