LIGAMAX-5MM ENDO CLIP APPLIER
Report
- Report Number
- 3005075853-2008-01775
- Event Type
- Malfunction
- Date Received
- September 18, 2008
- Date of Event
- August 12, 2008
- Report Date
- August 22, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
ADVANCER. EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RETURNED WITH THE TIP OF THE ADVANCER BROKEN. THE INSTRUMENT WAS CYCLED AND PEAR SHAPED THE REMAINING CLIPS DUE TO THE ADVANCER CONDITION. AS PER THE INSTRUCTIONS FOR USE: "PRIOR TO LOADING A CLIP IN THE JAWS, ENSURE THAT THE DEMARCATION BETWEEN THE JAWS AND THE INSTRUMENT SHAFT IS PAST THE END OF THE TROCAR CANNULA." FURTHER IN THE WARNING SECTION, THE IFU CONTAINS THE FOLLOWING WARNING, "DO NOT EXCESSIVELY APPLY A SIDE LOAD TO THE JAWS THAT WOULD CAUSE THEM TO PARTIALLY COLLAPSE AND POTENTIALLY RESULT IN A CLIP MALFORMATION. THE DEVICE JAWS SHOULD BE FULLY OPEN AND PARALLEL UPON INITIATING THE FIRING OF THE INSTRUMENT." WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAP HERNIA PROCEDURE, THE DEVICE MIS-FED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM ENDO CLIP APPLIER | FZP | ETHICON ENDO-SURGERY, LLC | NA | E4LA4H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |