FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 1171397 · Received September 22, 2008

Report

Report Number
2953200-2008-00845
Event Type
Injury
Date Received
September 22, 2008
Date of Event
August 27, 2008
Report Date
August 29, 2008
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVAL: RESULTS: VERY SHORT AORTIC NECK), (ENDOLEAK). CONCLUSION CODE - VERY SHORT AORTIC NECK. SECONDARY INTERVENTION.

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. THE PT HAS A VERY SHORT AORTIC NECK. IT WAS REPORTED THE FINAL ANGIOGRAM REVEALED A TYPE I ENDOLEAK. THE PHYSICIAN BELIEVES THAT IT MAY BE POSSIBLE THAT THE TYPE I ENDOLEAK WAS CAUSED BY MINIMAL FIXATION OF THE STENT GRAFT TO THE VESSEL WALL, DUE TO THE SHORT AORTIC NECK. THE PHYSICIAN TREATED THE PATIENT WITH ANOTHER MANUFACTURER'S STENT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention