COLLEAGUE CX VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-02840
- Event Type
- Malfunction
- Date Received
- September 24, 2008
- Date of Event
- January 4, 2007
- Report Date
- January 4, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 25 2007. EVALUATION SUMMARY: THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100 ML/HOUR, THE INFUSION PUMP FAILED THE ACCURACY TEST. THE SPECIFICATION OF THIS DEVICE IS +/-5%. INSPECTION OF THE DEVICE FOUND THAT THE ACCURACY FAILURE WAS CAUSED BY A FAULTY PUMP HEAD MECHANISM AND THEREFORE THE PUMP HEAD MECHANISM WAS REPLACED.
DURING SERVICE, THE PUMP HEAD MECHANISM FAILED THE ACCURACY TEST. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |