COLLEAGUE VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-02922
- Event Type
- Malfunction
- Date Received
- September 24, 2008
- Date of Event
- January 5, 2007
- Report Date
- January 5, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-7/20/05-015-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 25 2007. EVALUATION SUMMARY: THE REPORTED CONDITION OF FAIL CODE 703 WAS CONFIRMED. THIS FAILURE IS ASSOCIATED WITH THE USER INTERFACE MODULE COMMUNICATION WITH THE PUMP HEAD MODULE. THE USER INTERFACE MODULE WAS REPLACED. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER MDQ-CAPA-91. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE.
THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE THE BAXTER REPAIR TECHNICIAN REPORTED FAIL CODE 703. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INCIDENT INVOLVING THIS PUMP. NO ADDITIONAL INFORMATION IS AVAILABLE.