FDA Adverse Event Malfunction Summary report: N

CONCERTO

MDR report key: 1171356 · Received September 17, 2008

Report

Report Number
9611530-2008-00053
Event Type
Malfunction
Date Received
September 17, 2008
Date of Event
August 28, 2008
Report Date
September 16, 2008
Manufacturer
ARJO HOSPITAL EQUIPMENT AB
Product Code
FSA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN ARJO INVESTIGATOR HAS EXAMINED THE DEVICE AND FOUND MISSING SCREWS THROUGHOUT THE UNIT STRETCHER AND A SIGNIFICANT AMOUNT OF RUST. THERE WAS CRACKED FIBERGLASS THROUGHOUT THE STRETCHER AND A BROKEN LOCK ON THE RIGHT RAIL. THE UNIT COULD NOT BE FUNCTION TESTED DUE TO ITS CONDITION. THE UNIT IS COMPLETELY WORN. THE ROOT CAUSE FOR THE INCIDENT IS LACK OF MAINTENANCE. THE LOCKING HOOK THAT WAS BROKEN SHOULD BE REPLACED ONCE A YEAR; DEFECT PRODUCT WAS USED. THE PRODUCT IS 15 YEARS OLD AND, EVEN THOUGH NO EXPECTED LIFETIME IS STATED IN THE 1993 USER MANUAL, IT IS POINTED OUT THAT THE LIFETIME IS DEPENDING ON SUFFICIENT MAINTENANCE. IN THE CURRENT OPERATING AND PRODUCT MAINTENANCE INSTRUCTIONS MANUAL (OPI), IT IS STATED UNDER THE CHAPTER TITLED "SERVICE AND MAINTENANCE TO WEEKLY EXAMINE THE TROLLEY FOR DAMAGES. OBVIOUSLY THE PRODUCT WAS IN NEED FOR A SERVICE/TECHNICAL CHECK FOR QUITE SOME TIME. IN ITS CURRENT STATE, THE DEVICE IS DANGEROUS TO USE. THE MANUFACTURER STRONGLY RECOMMENDS IT BE REPLACED. AT A MINIMUM, WORN/DAMAGED PARTS HAVE TO BE REPLACED AND LIFT OPERATORS RETRAINED, ESPECIALLY REGARDING REGULAR INSPECTION AND MAINTENANCE OF THE EQUIPMENT.

Description of Event or Problem · 1

THE FACILITY REPORTS THE STAFF MEMBER WAS GIVING A SHOWER TO THE INJURED PARTY IN THE SHOWER TROLLEY. AFTER COMPLETING THE SHOWER, THE STAFF MEMBER TURNED THE RESIDENT ONTO THE STOMACH AND PROCEEDED WITH DRYING THE RESIDENT. THE SHOWER TROLLEY STRETCHER THEN CAME DETACHED FROM THE LEFT SIDE. THE RESIDENT FELL OUT OF THE SHOWER TROLLEY ON TO THE BATHROOM TILE FLOOR. THE RESIDENT SUSTAINED INJURIES TO THE SHOULDER AND ARM. THE COMPLETE INJURY DESCRIPTION WAS NOT PROVIDED. THE RESIDENT RECEIVED SOME MEDICAL TREATMENT AND WAS RELEASED THE SAME DAY FROM THE ER. MFR. (B) (4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO SHOWER TROLLEY FSA ARJO HOSPITAL EQUIPMENT AB BABXXXX

Patients

Seq Age Sex Outcome Treatment
1 Other