BELIMED STEAM STERILIZER MST-H TOP 5000
Report
- Report Number
- 3001061128-2021-00006
- Event Type
- Malfunction
- Date Received
- April 23, 2021
- Date of Event
- June 4, 2018
- Report Date
- April 22, 2021
- Manufacturer
- BELIMED AG
- Product Code
- FLE
- PMA / PMN Number
- K170228
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
DURING INITIAL INSTALLATION OF THE DEVICE AT A HEALTHCARE FACILITY, THE TECHNICIAN IN CHARGE INTENDED TO PROGRAM A PREVACUUM CYCLE BASED ON USER-SPECIFIED PARAMETERS. HE ENTERED THE PARAMETERS AND RENAMED THE PROGRAM TO IDENTIFY THE PREVACUUM CYCLE. HOWEVER, HE FAILED TO SAVE THE CHANGED PARAMETERS AND THE PROGRAM REVERTED TO THE ORIGINAL GRAVITY CYCLE PARAMETERS. THUS, WHENEVER THE SUPPOSEDLY PREVACUUM PROGRAM WAS RUN, A GRAVITY CYCLE WAS EXECUTED. THE PROGRAM WAS USED OCCASIONALLY FROM 1/11/2019 THROUGH 3/12/2021. SEVEN STERILIZERS WERE INSTALLED AT THE AFFECTED HEALTHCARE FACILITY AT THE TIME UNDER THE SUPERVISION OF THE SAME TECHNICIAN. FOUR OF THE STERILIZERS WERE AFFECTED BY A SIMILAR PROBLEM. THIS IS MEDICAL DEVICE REPORT NUMBER THREE OF FOUR. THE PROGRAMMING ERROR HAS BEEN CORRECTED AND THE PROGRAM IS RUNNING PROPERLY. CORRECTIVE AND PREVENTIVE ACTIONS HAVE BEEN INITIATED.
A USER-SPECIFIED PREVACUUM STERILIZATION CYCLE WAS INCORRECTLY PROGRAMMED BY THE TECHNICIAN DURING INITIAL INSTALLATION OF THE STERILIZER. THE PROGRAM RAN A GRAVITY CYCLE WITH ADDITIONAL STERILIZATION HOLD TIME. DEVICE STERILIZATION WAS ENSURED, BUT THE ERROR REPEATEDLY RESULTED IN WET PACKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 614508 | BELIMED STEAM STERILIZER MST-H TOP 5000 | STEAM STERILIZER | FLE | BELIMED AG | GR 9-6-15 HS2 | 25177 |