FDA Adverse Event Injury Summary report: N

TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET

MDR report key: 1171175 · Received September 22, 2008

Report

Report Number
3005099803-2008-04751
Event Type
Injury
Date Received
September 22, 2008
Date of Event
August 22, 2008
Report Date
August 26, 2008
Manufacturer
BOSTON SCIENTIFIC
Product Code
LQC
PMA / PMN Number
K040447
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOSCOPY (ERCP) PROCEDURE, A BASKET AND BASKET PULL-WIRE DETACHMENT AND SURGICAL INTERVENTION OCCURRED. THE LESION WAS LOCATED IN THE DISTAL COMMON BILE DUCT. THE RX LITHOTRIPTER COMPATABLE BASKET WAS ADVANCED TO THE STONE AND WAS ABLE TO SUCCESSFULLY CAPTURE THE STONE. UPON ATTEMPTING TO MANUALLY CRUSH THE STONE, THE STONE WAS UNABLE TO BE CRUSHED. THE PHYSICIAN PLACED THE HANDLE OF THE BASKET INTO A ANOTHER MFR'S PRESSURE GUN. UPON MECHANICAL CRUSHING OF THE STONE, THE DISTAL TIP OF THE BASKET FAILED TO DISENGAGE AND THE COIL OVER-SHEATH SEPARATED IN THE MID-CATHETER. THE HANDLE BROKE OFF WHICH IN TURN PULLED OFF THE OVER-SHEATH, LEAVING THE BASKET AND BASKET PULL-WIRE INSIDE THE PT. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE BASKET AND BASKET PULL WIRE BY ENLARGING A PREVIOUS SPHINCTEROTOMY WITH UNSPECIFIED BALLOON CATHETERS, HOWEVER, THIS WAS UNSUCCESSFUL. THE PT WAS TAKEN TO THE OPERATING ROOM WHERE THE BASKET AND BASKET PULL WIRE WERE SURGICALLY RETRIEVED. THE PT'S STATUS IS REPORTED AS "RECOVERING".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET LQC - LITHOTRIPSY BASKET LQC BOSTON SCIENTIFIC M00510880 0011841706

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention ALLIANCE PRESSURE GUN