CHLORAPREP SEPP
Report
- Report Number
- 3004932373-2021-00209
- Event Type
- Injury
- Date Received
- April 22, 2021
- Date of Event
- March 24, 2021
- Report Date
- July 30, 2021
- Manufacturer
- CAREFUSION 213, LLC 0113
- Product Code
- KXG
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
NO PHOTOS OR SAMPLES WERE AVAILABLE FOR EVALUATION. UNFORTUNATELY, AS A RESULT, BD WAS UNABLE TO VERIFY THE REPORTED ISSUE OR DEFINE A ROOT CAUSE. IF SAMPLES OR PHOTOS BECOME AVAILABLE, THIS RECORD WILL BE RE-OPENED AND RE-ASSESSED. PRODUCTION RECORD REVIEW WAS COMPLETED FOR BATCH/LOT 8264564 AND NO NON-CONFORMITIES, FAILURES, DEVIATIONS, OR REWORK ACTIVITIES OCCURRED DURING THE MANUFACTURING OF THIS LOT SIMILAR TO THE REPORTED ISSUE OR THAT COULD HAVE CONTRIBUTED TO THE REPORTED FAILURE MODE. NO FURTHER ACTION WILL BE TAKEN BASED ON NO ADVERSE TREND OBSERVED AND WILL CONTINUE TO BE TRACKED AND TRENDED.
IT WAS REPORTED THAT TWO MEMBERS OF THE STAFF HAVE HAD A SHARPS INJURY DUE TO THE GLASS BREAKING THROUGH THE PLASTIC. PER COMPLAINT FORM: I SPOKE TO HELEN ON THE PHONE THEY ARE IN RADIOLOGY AND USED THE CHLORAPREP SEPP THAT IS PART OF THE ROCIALLE CANNULATION PACK THAT IS USED IN WALES. THEY HAVE USED THE PRODUCT FOR A LONG TIME BUT TWO MEMBERS OF STAFF HAVE HAD A SHARPS INJURY DUE TO THE GLASS BREAKING THROUGH THE PLASTIC. ALSO BELOW PLEASE SEE THE INITIAL COMMUNICATION SENT TO ME VIA THE PRODUCT MANAGER. CONTACT INFO OMITTED. BUSINESS UNIT: INFECTION PREVENTION . PRODUCT LINE: CHLORAPREP. ENQUIRY: A COUPLE OF MY STAFF AND CUT THEIR FINGERS RECENTLY USING CHLORAPREP SEPP 0.67ML. THIS COULD BE A BATCH ISSUE? BATCH NUMBER 8264564.
(B)(4). INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).
IT WAS REPORTED THAT TWO MEMBERS OF THE STAFF HAVE HAD A SHARPS INJURY DUE TO THE GLASS BREAKING THROUGH THE PLASTIC. PER COMPLAINT FORM: I SPOKE TO (B)(6) ON THE PHONE THEY ARE IN RADIOLOGY AND USED THE CHLORAPREP SEPP THAT IS PART OF THE ROCIALLE CANNULATION PACK THAT IS USED IN WALES. THEY HAVE USED THE PRODUCT FOR A LONG TIME BUT TWO MEMBERS OF STAFF HAVE HAD A SHARPS INJURY DUE TO THE GLASS BREAKING THROUGH THE PLASTIC. ALSO BELOW PLEASE SEE THE INITIAL COMMUNICATION SENT TO ME VIA THE PRODUCT MANAGER. CONTACT INFO OMITTED. BUSINESS UNIT: INFECTION PREVENTION. PRODUCT LINE: CHLORAPREP. ENQUIRY: A COUPLE OF MY STAFF AND CUT THEIR FINGERS RECENTLY USING CHLORAPREP SEPP 0.67ML. THIS COULD BE A BATCH ISSUE? BATCH NUMBER 8264564.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 607323 | CHLORAPREP SEPP | 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL | KXG | CAREFUSION 213, LLC 0113 | 8264564 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR | Other |