COLLEAGUE 3 VOLUMETRIC INFUSION PUMP

MDR report key: 1171008 · Received September 24, 2008

Report

Report Number
6000001-2007-01512
Event Type
Malfunction
Date Received
September 24, 2008
Date of Event
December 1, 2006
Report Date
December 18, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 16 2007. THE REPORTED CONDITION OF FAILURE CODE 500:320 WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THE CAUSE OF THE REPORTED CONDITION WAS DUE TO A DEFECTIVE FRONT BEZEL. THE FRONT BEZEL WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE FACILITY REPORTED TO CUSTOMER SERVICE A PUMP WITH FAILURE CODE 500:320. IT IS UNKNOWN WHEN THIS FAILURE CODE OCCURRED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.