FDA Adverse Event Malfunction Summary report: N

LS CLAVE PORT EXT 7"

MDR report key: 1170830 · Received September 15, 2008

Report

Report Number
9615050-2008-00276
Event Type
Malfunction
Date Received
September 15, 2008
Date of Event
January 1, 2008
Report Date
August 25, 2008
Manufacturer
HOSPIRA DE COSTA RICA LTD.
Product Code
FPA
PMA / PMN Number
K915571
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS NOT RETURNED SINCE THIS IS A KNOWN ISSUE. THE ISSUE OF SPLIT OR BURST TUBING WHEN USING POWER INJECTORS WAS EVALUATED UNDER A FORMAL INVESTIGATION. IT WAS DETERMINED THAT THIS IS NOT A MANUFACTURING OR MATERIALS DEFECT. IT WAS COMMUNICATED TO THE CUSTOMER THAT STANDARD I.V. ADMINISTRATION SETS ARE INTENDED FOR GENERAL INFUSION, AND NOT RECOMMENDED FOR USE WITH POWER INJECTORS. IN ADDITION, THE INFUSION THERAPY ACCOUNT MANAGERS WERE ADVISED OF THIS ISSUE, AND WERE PROVIDED WITH A PRODUCT LIST OF THOSE HIGH-PRESSURE SETS THAT ARE APPROPRIATE FOR USE WITH POWER INJECTORS. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED A TUBING RUPTURE WHILE IN USE WITH A POWER INJECTOR; SUBSEQUENTLY, BLOOD LOSS WAS NOTED. THE TUBING SET WAS CONNECTED TO A POWER INJECTOR AND WAS BEING USED TO DELIVER AN UNSPECIFIED CONTRAST MEDIA AT A RATE OF 4ML/SEC TO 5ML/SEC. AFTER AN UNSPECIFIED LENGTH OF TIME IN USE, THE TUBING RUPTURED AT AN UNSPECIFIED LOCATION. THE CUSTOMER CONTACT REPORTED AN UNSPECIFIED VOLUME OF BLOOD WAS LOST. THE TUBING SET WAS REPLACED AND THE PROCEDURE WAS COMPLETED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. NO MEDICAL INTERVENTIONS WERE REQUIRED. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LS CLAVE PORT EXT 7" 80-FPA FPA HOSPIRA DE COSTA RICA LTD. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK UNSPECIFIED POWER INJECTOR| UNSPECIFIED CONTRAST MEDIA