FDA Adverse Event Malfunction Summary report: N

LARGE NEEDLE DRVIER ISNTRUMENT

MDR report key: 1170806 · Received September 12, 2008

Report

Report Number
2955842-2008-01269
Event Type
Malfunction
Date Received
September 12, 2008
Report Date
September 12, 2008
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
GCJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING PLACED THE INSTRUMENT ON A TEST SYSTEM AND THE INSTRUMENT WAS INTERMITTENTLY RECOGNIZED. IT WAS DETERMINED THAT THE RECOGNITION ISSUE EXPERIENCED BY THE CUSTOMER WAS DUE TO THE PROGRAMMING ERROR OR A FAULTY RTI BOARD. ENGINEERING ALSO OBSERVED THE DISTAL END OF THE MAIN TUBE HAS A COUPLE OF SECTIONS WITH MATERIAL REMOVED DUE TO DEEP SCRATCHES. THE SCRATCHES ARE UNDER .250" LONG AND NOT ALIGNED WITH THE TUBE AXIS. BASED ON THE LOCATION AND APPEARANCE, THE DAMAGE WAS MOST LIKELY CAUSED BY INSTRUMENT COLLISIONS AND/OR ROUGH HANDLING. NO OTHER DAMAGE FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS. HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S PROSTATECTOMY PROCEDURE THE LARGE NEEDLE DRIVER INSTRUMENT WAS NOT RECOGNIZED BY THE SYSTEM. NO ADDITIONAL INFO WAS PROVIDED. NO PT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LARGE NEEDLE DRVIER ISNTRUMENT ENDOSCOPIC INSTRUMENT GCJ INTUITIVE SURGICAL, INC 420006-05 081242 390

Patients

Seq Age Sex Outcome Treatment
1 ACCESSORIES| DA VINCI S SURGICAL SYSTEM