FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 1170804 · Received September 12, 2008

Report

Report Number
2955842-2008-01267
Event Type
Malfunction
Date Received
September 12, 2008
Report Date
September 12, 2008
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONDUCTED PERFORMANCE TESTING AND FOUND THE INSTRUMENT TO BE SUCCESSFULLY RECOGNIZED BY A DA VINCI S SYSTEM. THE POGO PINS DO NOT STICK AND ARE NOT CONTAMINATED. ENGINEERING COULD NOT REPLICATE THE CUSTOMER REPORTED RECOGNITION ISSUE. ENGINEERING ALSO OBSERVED THE DISTAL END OF THE MAIN TUBE TO HAVE VARIOUS SCRATCH MARKS CONCENTRATED IN A 1.6" LONG SECTION ON ONE SIDE OF THE TUBE. A FEW SCRATCHES ARE DEEP ENOUGH TO HAVE REMOVED TUBE MATERIAL. BASED ON THE LOCATION AND APPEARANCE, THE DAMAGE WAS MOST LIKELY CAUSED BY INSTRUMENT COLLISIONS AND OR ROUGH HANDLING. NO OTHER DAMAGE FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS. HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS NOT RECOGNIZED BY THE DA VINCI S SYSTEM. NO ADDITIONAL INFO WAS PROVIDED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ELECTROSURGICAL INSTRUMENT GEI INTUITIVE SURGICAL, INC 420205-03 1904071 377

Patients

Seq Age Sex Outcome Treatment
1 ACCESSORIES| DA VINCI S SURGICAL SYSTEM| AN ELECTROSURGICAL UNIT