COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-02768
- Event Type
- Malfunction
- Date Received
- September 24, 2008
- Date of Event
- January 4, 2007
- Report Date
- January 4, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 24 2007. EVALUATION SUMMARY:THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT ML/HOUR, THE CHANNEL A PUMP HEAD MECHANISM FAILED THE ACCURACY TEST. THE SPECIFICATION OF THIS DEVICE IS +/-5%. INSPECTION OF THE DEVICE FOUND THAT THE ACCURACY FAILURE WAS CAUSED BY A FAULTY PUMP HEAD MECHANISM, AND THEREFORE THE CHANNEL A PUMP HEAD MECHANISM WAS REPLACED.
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, THE CHANNEL A PUMP HEAD MECHANISM FAILED THE ACCURACY TEST. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.