FDA Adverse Event Death Summary report: N

QUICK PACE

MDR report key: 11707 · Received January 25, 1994

Report

Report Number
11707
Event Type
Death
Date Received
January 25, 1994
Date of Event
December 11, 1993
Report Date
December 21, 1993
Manufacturer
PHYSIO CONTROL CORP.
Product Code
DRO
Report Source
User Facility report
Reporter Location
CA, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

AN EXTERNAL PACER FAILED WHEN APPLIED DURING A CARDIAC ARREST. THE MACHINE READ "PACING LEADS NOT ON." UNABLE TO EXTERNALLY PACE. PATIENT EXPIRED. LATER A NEW CABLE WAS ATTACHED AND THE QUICK PACE PERFORMED PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUICK PACE EXTERNAL PACER DRO PHYSIO CONTROL CORP. 804641-01

Patients

Seq Age Sex Outcome Treatment
1 39 YR Death