INSPECTRA STO2

MDR report key: 1170604 · Received September 16, 2008

Report

Report Number
2184064-2008-00001
Event Type
Malfunction
Date Received
September 16, 2008
Manufacturer
HUTCHINSON TECHNOLOGY, INC.
Product Code
MUD
PMA / PMN Number
R061619
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE OVER TEMPERATURE ERROR MESSAGE FUNCTIONED AS INTENDED TO ALERT THE USER OF THE OVER TEMPERATURE CONDITION. CONCLUSIONS: A ROOT CAUSE INVESTIGATION WAS INITIATED ON (B) (6) 2008, FOR THE POWER LINE FILTER MALFUNCTION. THE 8051 PROCESSOR WITHIN THE DEVICE DETECTED THE HEAT FROM THE POWER LINE FILTER MALFUNCTION AND COMMUNICATED AN OVER TEMPERATURE ERROR MESSAGE, WHICH WAS DISPLAYED ON THE DEVICE SCREEN.

Description of Event or Problem · 1

A HEALTHCARE PROFESSIONAL OBSERVED A SYSTEM TEMPERATURE ERROR MESSAGE ON THE MONITOR SCREEN. MONITOR WAS IN THE ICU ATTACHED TO AN IV POLE AND PLUGGED INTO AN ELECTRICAL OUTLET DURING THE TIME THE OVER TEMPERATURE MESSAGE WAS OBSERVED. UPON EVALUATION OF THE RETURNED PRODUCT, OVERHEATING ISOLATED TO THE POWER LINE FILTER WAS OBSERVED. THE POWER LINE FILTER IS CONTAINED WITHIN AN ENCLOSURE WITH A UL FLAMMABILITY RATING OF 94V-0. NO INJURIES WERE REPORTED.