FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER, 45 MM STAPLE LINE - 4

MDR report key: 1170355 · Received September 12, 2008

Report

Report Number
3005075853-2008-01666
Event Type
Malfunction
Date Received
September 12, 2008
Date of Event
July 24, 2008
Report Date
July 26, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DAMAGED FIRING TRIGGER TEETH. EVALUATION SUMMARY: THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A PARTIALLY RELOAD ON THE DEVICE. THE CARTRIDGE LOCK OUT TAB WAS BENT. THE DAMAGE TO THE CARTRIDGE LOCKOUT TAB IS CONSISTENT WITH DAMAGE OBSERVED WHEN THE FIRING CYCLE IS STARTED, INTERRUPTED, RELEASED, AND RESTARTED. WHEN FIRING THE DEVICE, MAKE SURE THAT THE FIRING STROKE IS COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR MORE INFORMATION. THE RETURNED DEVICE WAS FOUND TO BE NON FUNCTIONAL, AS THE FIRING MECHANISM WAS NOTED TO BE DAMAGED. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE COULD NOT CUT THE TISSUE. THE SURGEON CHANGED TO ANOTHER RELOAD, BUT IT STILL COULD NOT CUT THE TISSUE. THE SURGEON USED ANOTHER SAME LIKE DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER, 45 MM STAPLE LINE - 4 GDW ETHICON ENDO-SURGERY, LLC. NA E4K47Y

Patients

Seq Age Sex Outcome Treatment
1