UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEMS
Report
- Report Number
- 2050012-2008-00052
- Event Type
- Malfunction
- Date Received
- September 11, 2008
- Date of Event
- August 20, 2008
- Report Date
- September 11, 2008
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NO INFORMATION
Narratives
QC WAS RUN ON THE DAY OF THE EVENT AND RESULTS WERE WITHIN THE LAB'S ESTABLISHED RANGES. LATER IN THE DAY, THE LAB WAS NOTIFIED THAT K RESULTS WERE LOW. QC WAS RUN AGAIN AND RECOVERED LOW. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S LAB. THE ISE SYSTEM WAS CULTURED AND FOUND TO BE CONTAMINATED WITH PSEUDOMONAS. THE FSE DECONTAMINATED THE ISE SYSTEM, REPLACED THE K TIP AND CARBON BRIDGE. AS OF (B) (4) 2008, NO FURTHER LOW K RESULTS WERE GENERATED. A CLEAR ROOT CAUSE HAS NOT BEEN DETERMINED FOR THIS EVENT. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT. (B) (4).
A CUSTOMER CONTACTED BECKMAN COULTER INC (BCI) REGARDING ERRONEOUSLY LOW POTASSIUM (K) RESULTS GENERATED BY THE UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEMS FOR MULTIPLE PTS. ERRONEOUSLY LOW K RESULTS WERE REPORTED OUT OF THE LAB FOR SEVENTEEN (17) PTS. THE RESULTS WERE IN THE RANGE OF 3.0-4.9 MMOL/L (B) (4). CUSTOMER RECALIBRATED THE ION SELECTIVE ELECTRODE (ISE) SYSTEM, RAN QC WITH ACCEPTABLE RESULTS AND RE-TESTED THE SAMPLES. REPEATED K RESULTS WERE HIGHER AND 17 AMENDED REPORTS WERE ISSUED. (B) (4). THE CUSTOMER STATED TREATMENT WAS NOT INITIATED OR WITHHELD ON ANY PTS BASED ON THE LOW K RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEMS | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |