COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-02779
- Event Type
- Malfunction
- Date Received
- September 24, 2008
- Date of Event
- January 10, 2007
- Report Date
- January 10, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 24 2007. EVALUATION SUMMARY: THE CONDITION OF FAILURE CODE 570 WAS CONFIRMED. THIS FAIL CODE WAS CAUSED BY DEPLETED BATTERIES. THE BATTERIES WERE REPLACED. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE.
THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE THE BAXTER REPAIR TECHNICIAN NOTED FAIL CODE 570 IN THE EVENT HISTORY. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INCIDENT INVOLVING THIS PUMP. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |